Harmonic and Flicker Generator 01
The Harmonics & Flicker Generator (HFG01) has聽been designed for the purpose of verifying harmonic聽and flicker test equipment.
The HFG01 provides an easy and聽reliable way to externally check the performance of聽the measurement system to the IEC/EN 61000-3-2聽harmonics and IEC/EN 61000-3-3 flicker standards;聽particularly important as these tests rely on software聽control and calculation, and for which there is no聽intuitive sense of the response.
The HFG01 provides a series of harmonic and flicker聽disturbances of a nominal but stable level. This allows the聽user to periodically verify their test equipment, helping聽maintain compliance with standards and laboratory quality聽procedures. Alternatively, due to its stability, it may be used聽as a transfer standard from a known, calibrated test system.
The HFG01 is a standalone device and requires no additional聽equipment. It connects directly to the test equipment and聽simulates the equipment under test (EUT), generating known,聽repeatable levels of harmonic and flicker disturbance.
Key Features
- Stable load simulation
- Injects harmonics to IEC/EN 61000-3-2 and Flicker to IEC/EN 61000-3-3
- Compact and portable
- Independent power supply
Features/Applications
Features
Stable聽load simulation
- Repeatable measurements for test system verification
Injects harmonics to IEC/EN 61000-3-2 and聽flicker聽to IEC / EN 61000-3-3
- Evaluation of test systems specifically to聽EN standards
Harmonic test modes
- Steady-state harmonic-rich load current,聽representing a fixed load
- Harmonic-rich load currents fluctuating between聽two load conditions
Flicker test modes
- Fixed level of mains disturbance at 1 Hz rate
- Fixed level of mains disturbance at 8.33 Hz rate
Compact and portable
- Comparisons between sites and environments
Independent power supply
- Separate power supply to the HFG01, to avoid contamination of results
Applications
- Harmonics and flicker measurement systems validation and verification
- Reference source for
- Daily pre-test verification checks if required by聽the accreditation authorities
- Long-term performance monitoring
- Comparison of different harmonics and flicker measurement systems
Specifications
Test supply connector
- 1.3 m captive lead with CEE 7/7 mains plug for connection to test equipment
Dimensions
- 330 mm 脳 320 mm 脳 170 mm
Weight
- 6.5 kg
Power Supply
- External power supply, 24 Vdc, 1A maximum
Indicators
- Thermal shutdown
Harmonic current
- Up to 40th聽harmonic: 50 Hz to 2 kHz
Flicker disturbance
- 1 Hz and 8.33 Hz
Kits
Standard Order Kits
HFG01KIT01 鈥 Standard HFG01 harmonics and flicker generator聽kit
Parts included
- HFG01 harmonic and flicker generator
- CAL12 鈥 measurement of harmonics and聽flicker generated, all modes
- External power supply
- User Manual
Manufacturer's Calibrations
CAL12 鈥 Measurement of harmonic and flicker disturbance generated
- Harmonics
- Measurement of load current made according to IEC / EN 61000-3-2 in Steady State and Fluctuating Harmonics modes. Fundamental (50 Hz) to 40th harmonic.
- Flicker
- Measurement of short term flicker (Pst) made according to IEC / EN 61000-3-3 with聽disturbance at 1 Hz and 8.33 Hz rates